Catalog No: RE1600H
Human NP(Nandrolone Phenylpropionate) ELISA Kit
Copy Product Information

48T

$299.00

96T

$399.00

Product Details

Species:

Human

Alternative Name:

NPP, Nandrolone phenpropionate, 19-Nortestosterone phenylpropionate, Nandrolone hydrocinnamate, 19-Nortestosterone 17β-phenylpropionate

Assay Type:

Competitive

unit:

pg/mL

Standard:

2000pg/mL

Sensitivity:

18.75 pg/mL

Detection Range:

31.25-2000pg/mL

Assay Time:

1.5H

Sample Type:

Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50μL

Research:

Signal transduction;

Test Principle:

This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Human NP. During the reaction, Human NP in the sample or standard competes with a fixed amount of Human NP on the solid phase supporter for sites on the Biotinylated Detection Ab specific to Human NP. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of Human NP in tested samples can be calculated by comparing the OD of the samples to the standard curve.

Technical Data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
pg/mL OD Corrected
2000 0.19
1000 0.326
500 0.541
250 0.875
125 1.303
62.5 1.695
31.25 2.071
0 2.511

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.

Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean(ng/mL) 85.1 401.06 543.17 33.93 282.71 833.61
Standard deviation 2.28 19.24 39.17 3.71 24.07 42.87
C V (%) 4.63 4.77 3 5.34 4.61 5.02

Rate of recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type Range(%) Average Recovery(%)
Serum (n=8) 93-101 97
EDTA plasma (n=8) 85-103 94
Cell culture media (n=8) 95-105 100

Linear

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) EDTA plasma (n=5) Cell culture media (n=5)
1:2 Range (%) 80-94 85-92 90-99
Average (%) 91 86 91
1:4 Range (%) 87-96 85-92 95-103
Average (%) 95 89 97
1:8 Range (%) 85-98 80-96 86-92
Average (%) 97 96 88
1:16 Range (%) 95-104 93-105 91-102
Average (%) 96 103 98

Assay Procedures

Description